Table 1 Laboratory Pima™ testing using venous blood

From: Performance evaluation of the Pima™ point-of-care CD4 analyser using capillary blood sampling in field tests in South Africa

Phase I

EDTA venous blood

Pima #1

Pima #2

Pima #3

Pima #4

All

Pima DATA

N

53

61

59

58

100**

Range of CD4 counts*

2-1297

4-1217

5-1190

7-972

4.5-1235

Mean CD4 count*

349

334

354

317

318

%Similarity to Predicate

( %SIM Mean ± SD)

99.26 ± 5.34

99.15 ± 5.35

99.64 ± 6.42

97.17 ± 5.01

98.17 ± 4.57

%SIM CV #

5.38%

5.40%

6.44%

5.16%

4.66%

BA Bias* ± 1 STDev

(PIMA - PLG)

(95% CI of bias mean)

-14.38 ± 37.9

(-23.8 to -3.9)

-11.05 ± 42.4

(-21.9 to 0.19)

-13.6 ± 53.3

(-27.5 to 0.22)

-22.6 ± 40.3

(-33.2 to -12.0)

-17.3 ± 36.7

(-24.0 to -10.03)

BA 95% LOA*

-88.7 to 59.9

-94.1 to 72.1

-118.2 to 90.8

-101.6 to 56.4

-89.4 to 54.7

  1. Statistical analyses across four Pima™ analysers tested in the laboratory during Phase 1 using a fixed 20 mm3 volume of manually pipetted EDTA venous blood and compared against predicate flow cytometry (PLG/CD4). All values, with the exception of N, refer to CD4 cell counts in cells/mm3. BA = Bland-Altman statistical analysis. LOA = limits of agreement. #Precision of Pima™ to Predicate method expressed as %SIM CV.
  2. *cells/mm3
  3. **The average Pima/CD4 value was used for comparison to predicate CD4 where more than two simultaneous Pima values were obtained.